HSS 1026 Hydro Surveillance System

The INVALCO HSS 1026 oil in Water monitor has been designed for municipal and industrial applications to measure PPM levels of hydrocarbons in aqueous solutions. Typical applications include PPM trace amounts of oil in effluent water from storm water runoff, oil in cooling water, produced water, and oil/water separators. Other measurements and mediums can be monitored on request (i. e. colorants in fluids, etc.).

A continuous sample is directed into the cuvette using a pumped or process pressure source. A minimal flow rate is required. The sample is released from the cuvette by a gravity flow to a drain or sump.

The INVALCO HSS 1026 uses a light scatter technique to respond to oil contamination in the water. A slip stream approach directs a continuous sample flow through the Hydro Surveillance System and back into the process stream. While it passes through the sample cell a controlled light source is directed into the water. The emulsified oils in the water will scatter the light toward the light sensors placed strategically around the cell. The intensity of light energy is measured to provide an indication of the ppm concentration.

Overview

Features and Benefits

  • Ultrasonic disc generates a continuous cleaning action within the sample cell to reduce maintenance frequency
  • Compensation for temperature and lamp deterioration minimizes re-calibration requirements
  • Long life lamp
  • Desiccant chamber keeps electronics dry in humid conditions
  • Continuous display updates every one second
  • No consumables or chemical used
  • Sample flow returns to the process
  • Sample cell can be exchanged with prepared
  • Samples for easy testing and calibration
  • No tools necessary for routine maintenance

Specifications

Measurement Range0-100 PPM
Accuracy ±5% of span
Resolution 0.1 ppm
Response TimeAdjustable
Display Multi-Line Liquid Crystal Backlit Display
Alarms Two Programmable, 120-240VAC 2A Form C Relay
Analog OutputPowered 4-20mA, 600 – drive
Communications PortBi-directional RS-485, Modbus Optional
Maximum Water PressureIntegral pressure regulator rated 1380kPa (200 PSI.) Also refer to Flow Rate
Flow Rate100 ml/min. – 1 liter/min. (. 026-. 26 Gal/min)
Operating Temperature1°C – 50°C (34°F – 122°F)
Wetted MaterialsNylon, Borosilicate Glass, Silicon, Polypropylene, Stainless Steel
Sample Temperature Range1°C – 50°C (34°F – 122°F)
Power Supply Ordering Information100 – 240 VAC, 47 – 63 Hz, 80VA
Insulation RatingDouble Insulated, Pollution Degree 2, Overvoltage Category II
Environmental ConditionsNot recommended for outdoor use. Altitude up to 2000 meters. Up to 95 % RH (non-condensing)
Enclosure Rating IP 66 /NEMA 4X
Certifications CE Approved, ETL listed to UL 3111-1 & ETL Certified to CSA 22.2 No. 1010-1-92
Shipping Weight 2.5 kg (5.5 lbs.)
Warranty 1 Year from date of shipment

Please consult factory for Part Number and additional information.

Downloads

Request
More Information

We welcome and value your feedback please feel free to contact us at the information below. We look forward to assisting you.